Determination of phosphate by carbon paste electrode modified with SMZ by potentiometric method
عنوان مقاله: Determination of phosphate by carbon paste electrode modified with SMZ by potentiometric method
شناسه ملی مقاله: ZEOLITE01_025
منتشر شده در همایش بین المللی ژئولیت ایران در سال 1387
شناسه ملی مقاله: ZEOLITE01_025
منتشر شده در همایش بین المللی ژئولیت ایران در سال 1387
مشخصات نویسندگان مقاله:
خلاصه مقاله:
The utility of carbon paste electrode modified with SMZ (Na-Y zeolite modified with HDTMA
surfactant) for the potentiometric determination of phosphate in aqueous solution is demonstrated. The
electrode exhibits linear response to phosphate over a wide concentration range (1.58 × 10-5 – 1.00 × 10-2 M)
with Nernstian slope of 30 ± 2 mV per decade in temperature range 25 to 50 oC . It has a response time of
about 15 s and can be used for a period of two months with good reproducibility and repeatability. The
detection limit of this electrode was 1.5 × 10-5 M. The electrode response for phosphate determination in pH
range 6-12, was constant and independent from pH. This chemically modified carbon paste electrode was
successfully used for the determination of phosphate in ammonium phosphate chemically fertilizer sample.
کلمات کلیدی: SMZ, Potentiometry, Phosphate
صفحه اختصاصی مقاله و دریافت فایل کامل: https://civilica.com/doc/66075/