Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 14، شماره: 3
سال انتشار: 1396
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 355
فایل این مقاله در 7 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-14-3_002
تاریخ نمایه سازی: 18 تیر 1398
چکیده مقاله:
Introduction: Coronary angiography is the most common angiographic procedure for diagnosis and treatment of the heart diseases. Herein, we aimed to evaluate the entrance surface dose (ESD), dose area product (DAP), as well as cancer risk in interventional cardiology procedures. Materials and Methods: This study was conducted during July-December 2015 at Shahid Madani Heart Center in Khorramabad, Iran. A total of 225 adult patients including 122 females and 103 males regardless of the risk factors for coronary diseases were participated. Of them, 199 and 26 patients underwent diagnostic coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA), respectively. Each patient underwent CA or PTCA separately. All the procedures were carried out using Siemens angiography system with the pulsed fluoroscopy of 10-30 pulses/s and cine frame rate of 15 frames/s. DAP, ESD, fluoroscopy time (FT), as well as the number of sequences and frames per sequence were collected for each 199 CA and 26 PTCA procedures. Results: The median values of DAP were 19.77±14.88 and 57.11±33.36 Gy.cm2 in CA and PTCA, respectively. In addition, the median values of ESD were 323.12±245.39 and 1145.22±594.42 mGy in CA and PTCA, respectively. FTs were 114.59±74.33 s in CA and 424.15±292.93 s in PTCA. Conclusion: The average patient dose and cancer risk estimates in both CA and PTCA were consistent with the reference levels. However, in agreement with other interventional procedures, dose levels in the interventional cardiology are influenced by staff and clinical protocols, as well as the type of equipment.
کلیدواژه ها:
نویسندگان
Mehrdad Gholami
Department of Medical Physics & Radiation Sciences, Lorestan University of Medical Sciences, Khorram Abad, Iran
Somayeh gharloghi
Department of Medical physics and Radiation Sciences, School of Para Medicine, Lorestan University of Medical Sciences, Khorramabad, Iran
Soodabeh Zare
Department of Public Health, Lorestan University of Medical Sciences, Khorramabad, Iran
Arezoo Saki
Student Research Committee, Lorestan University of Medical Sciences, Khorramabad, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :