Telecobalt Machine Beam Intensity Modulation with Aluminium Compensating Filter Using Missing Tissue Approach
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 15، شماره: 1
سال انتشار: 1397
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 222
فایل این مقاله در 14 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-15-1_007
تاریخ نمایه سازی: 18 تیر 1398
چکیده مقاله:
Introduction: The present study aimed to generate intensity-modulated beams with Aluminium compensating filters for a conventional telecobalt machine based on the outputs of a treatment planning system (TPS) performing forward planning and cannot simulate directly the compensating filter.
Materials and Methods: In order to achieve the beam intensity modulation during treatment planning with the TPS, we used a bolus placed on the surface of a tissue-equivalent phantom. The treatment plans replicated on the telecobalt machine with the bolus were represented with compensating filters placed at a certain distance from the phantom surface. An equation was proposed for the conversion of the bolus thickness to the compensating filter thickness such that any point within the phantom would receive the planned dose. Correction factors were introduced into the proposed equation to account for the influences of field size, treatment depth, and applied bolus thickness. The proposed equation was obtained based on the analyses of empirical data measured in a full scatter water phantom with and without the compensating filter.
Results: According to the results, the dosimetric verification of the proposed approach outputs in a solid water phantom with calibrated Gafchromic EBT2 films were comparable to that of the TPS with deviation of ±4.73% (mean: 2.98±1.05%).
Conclusion: As the findings of the present study indicated, the discrepancy between the measured doses and TPS-estimated doses was within the tolerance of ±5%, which is recommended for dose delivery in external beam radiotherapy. Therefore, the proposed approach is recommended for clinical application.
کلیدواژه ها:
نویسندگان
Samuel Tagoe
National Centre for Radiotherapy & Nuclear Medicine, Korle Bu Teaching Hospital, Accra, Ghana; Physics Department, University of Cape Coast, Ghana; and School of Biomedical and Allied Health Sciences, University of Ghana, Accra,
Samuel Mensah
Department of Physics, School of Physical Sciences, University of Cape Coast, Cape Coast, Ghana.
John Justice Fletcher
Department of Applied Physics, University for Development Studies, Navrongo Campus, Navrongo, Ghana.
Evans Sasu
National Centre for Radiotherapy and Nuclear Medicine, Korle Bu Teaching Hospital, Accra, Ghana.
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :