Qualitative evaluation of Chang method of attenuation correction on heart SPECT by using custom made heart phantom

سال انتشار: 1382
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 408

فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_IRJNM-11-2_002

تاریخ نمایه سازی: 27 فروردین 1399

چکیده مقاله:

SPECT detects γ-rays from administered radiopharmaceutical within the patient body. γ-rays pass through different tissues before reaching detectors and are attenuated. Attenuation can cause artifacts; therefore different methods are used to minimize attenuation effects. In our study efficacy of Chang method was evaluated for attenuation purpose, using a custom made heart phantom. Due to different tissues surrounding heart, attenuation is not uniform more over activity distribution around heart is also non-uniform. In Chang method, distribution of radioactivity and attenuation due to the surrounding tissue is considered uniform. Our phantom is a piece of plastic producing similar SPECT image as left ventricle. A dual head, ADAC system was used in our study. Images were taken by 180° (Limited angle) and 360° (Total rotation). Images are compared with and without attenuation correction. Our results indicate that Chang attenuation correction method is not capable of eliminating attenuation artifact completely in particular attenuation effects caused by breast.

نویسندگان

Abbas Takavar

Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran

Mohammad Eftekhari

Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran

Davood Beiki

Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran

Mohsen Saghari

Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran