New approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras

سال انتشار: 1396
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 288

فایل این مقاله در 9 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_IRJNM-25-1_005

تاریخ نمایه سازی: 21 بهمن 1398

چکیده مقاله:

Introduction:SURGEOSIGHT is a dedicated intra-operative hand-held gamma camera designed and manufactured by our team to be used for lymphoscintigraphy. Although conventional gamma cameras are widely used, they lack the capability of the sentinel lymph node (SLN) imaging in the operation room. Like conventional gamma cameras it is necessary to calibrate SURGEOSIGHT for linearity distortion, energy and flood correction. Our main objective in this study is to develop and implement three techniques for linearity, energy and uniformity corrections. Methods: The camera is made up of a pixelated cesium iodide (sodium-activated) (CsI(Na)) scintillation crystal with 1 × 1 mm2 element size along with  a Hamamatsu H8500 flat-panel multi-anode (64 channels) photomultiplier tube. All triple corrections required lookup tables (LUTs). To generate all three LUTs, one uncalibrated acquisition is necessary. Linearity distortion correction was based on localization of physical pixels of crystal and thin plate spline interpolation to estimate the amount of distortion. Energy correction LUT was simply a photo-peak map of incoming events which was estimated using self-organizing map technique. Results: The results showed improvements qualitatively and quantitatively. The extrinsic energy resolution was enhanced from 50% to about 20.6%. Integral uniformity and the differential uniformity, after uniformity correction, in useful field-of-view (UFOV) was measured 9.5% and 4.5%, respectively. Conclusion: Results substantiated that the correction techniques guarantee uniform and accurate output of the SURGEOSIGHT which is desirable for intra-operative localization of the sentinel lymph nodes in breast cancer and also for imaging of other superficial tumors.

نویسندگان

Afshin Akbarzadeh

Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran|Radiation Research Center, Faculty of Paramedicine, AJA University of Medical Sciences, Tehran, Iran

Valiallah Saba

Radiation Research Center, Faculty of Paramedicine, AJA University of Medical Sciences, Tehran, Iran

Mohammad Reza Ay

Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran

مراجع و منابع این مقاله:

لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :
  • Bischof Delaloye A. Progress in the diagnosis and treatment of ...
  • Nishimura T. Angina pectoris: progress in diagnosis and treatment. II. ...
  • Anger HO. Scintillation camera. Rev. Sci. Instrum. 1958;29(1):27-33. ...
  • Gray RM, Macovski A. Maximum a posteriori estimation of position ...
  • Barrett HH, Hunter WC, Miller BW, Moore SK, Chen Y, ...
  • Bruyndonckx P, Leonard S, Tavernier S, Lemaitre C, Devroede O, ...
  • Jeong MH, Choi Y, Chung YH, Song TY, Jung JH, ...
  • Jeong MH, Choi Y, Chung YH, Song TY, Jung JH, ...
  • Hung N, Joung J, Lee K, Kim Y. Development of ...
  • Inbar D, Gafni G, Grimberg E, Koren J. Gamma camera ...
  • Berlad G, Wainer N. Gamma camera with two sequential correction ...
  • España S, Deprez K, Van Holen R, Vandenberghe S. Fast ...
  • Thanasas D, Maintas D, Georgiou E, Giokaris N, Karabarbounis A, ...
  • Knoll GF, Strange DR, Bennett MC. Radiation signal processing system. ...
  • Muehllehner G, Colsher JG, Stoub EW. Correction for field nonuniformity ...
  • Stoub EW, Colsher JG, Muehllehner G. Distortion correction method and ...
  • Arseneau RE. Non-uniformity energy correction method and apparatus. Patent US4323977 ...
  • Arseneau RE. Dynamic threshold for scintillation camera. Patent US4475042 A. ...
  • Barfod JM. Gamma camera with image uniformity by energy correction ...
  • Spector SS, Brookeman VA, Kylstra CD, Diaz NJ. Analysis and ...
  • Muehllehner G, Luig H. Septal penetration in scintillation camera collimators. ...
  • Muehllehner G. Radiation imaging device. Patent US3745345 A. 1973. ...
  • Bertelsen H. Gamma camera with automatc adjustment of the energy ...
  • Richey JB, Wake RH, Wilson HH, Cheiky MC. Dynamic uniform ...
  • Tararine M. Method for correcting the uniformity of a gamma ...
  • Kaviani S, Zeraatkar N, Sajedi S, Akbarzadeh A, Gorjizadeh N, ...
  • Ay M, Arabi H, Farahani M, Zeraatkar N, Sarkar S, ...
  • Gorjizadeh N, Zeraatkar N, Kaviani A, Far SF, Farahani M, ...
  • Grau V, Mewes AU, Alcañiz M, Kikinis R, Warfield SK. ...
  • Bookstein FL. Principal warps: thin-plate splines and the decomposition of ...
  • Kohonen T. The self-organizing map. Proc IEEE. 1990;78(9):1464-80. ...
  • Cottrell M, Olteanu M, Rossi F, Villa-Vialaneix N. Theoretical and ...
  • de Bodt E, Cottrell M, Verleysen M. Statistical tools to ...
  • Erwin E, Obermayer K, Schulten K. Self-organizing maps: ordering, convergence ...
  • Tolat VV. An analysis of Kohonen s self-organizing maps using ...
  • Palomares SE. Calibration of monolithic crystal-based detectors. Patent US20150276953 A1. ...
  • National Electrical Manufacturers Association. NEMA standards publication NU 1-2007: Performance ...
  • Levin CS, Hoffman EJ, Tomai MP, MacDonald LR. PSPMT and ...
  • David S, Georgiou M, Fysikopoulos E, Belcari N, Loudos G. ...
  • Deprez K, Van Holen R, Vandenberghe S. A high resolution ...
  • Bouckaert C, Marcinkowski R, Deprez K, Espana Palomares S, Van ...
  • Ferretti A, Chondrogiannis S, Marcolongo A, Rubello D. Phantom study ...
  • نمایش کامل مراجع