Application of Hydroxyapatite on Ti Via Cyclic Voltammetry Technique and Evaluating its Corrosion Behavior in Artificial Saliva

سال انتشار: 1398
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 44

فایل این مقاله در 10 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_ICAC-9-32_002

تاریخ نمایه سازی: 5 آذر 1402

چکیده مقاله:

Hydroxyapatite (HA, Ca۱۰(PO۴)۶(OH)۲) has been widely used as a coating material for dental implants for many years, due to its close similarity of chemical composition with bone material and its ability to bond with bone.  In this study, a hydroxyapatite coating was formed on Ti substrate, employing the electrochemical coating method of cyclic voltammetry, in order to enhance the corrosion resistance. Formation of Hydroxyapatite on Ti is confirmed by the XRD and SEM analyses. Polarization curves revealed a substantial improvement in the corrosion potential from -۲۸۹ to -۱۹۳ mV (Ag/AgCl) and decrease in corrosion-current density, for the hydroxyapatite coatings in artificial saliva compared with pure titanium that showed positive role of HA on Ti corrosion behavior